Noninvasive estimation of aortic pressure waveform based on simplified Kalman filter and dual peripheral artery pressure waveforms

Computer Methods and Programs in Biomedicine(2022)

引用 1|浏览15
暂无评分
摘要
•The article proposes a novel SKF approach for reconstructing the central aortic pressure waveform from dual peripheral artery pressure waveforms.•The main aim is to establish a cross-relation with the peripheral pressure measured noninvasively, and to use SKF's iterative optimization approach to derive an adaptive transfer function.•The adaptive transfer function provides a patient-specific method to meeting the demand for a clinically useful way to reliably estimate central aortic pressure from non-invasive peripheral pressure measurements.•The performance of the proposed method is superior to that of other state-of-the-art methods.
更多
查看译文
关键词
Aortic pressure,Peripheral artery pressure,Simplified Kalman filter,Canonical correlation analysis,Signal-to-noise ratio,Noise-tolerance
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要