Um insucesso de desfibrilhação por CDI resolvido de uma forma pouco habitual

Revista Portuguesa De Pneumologia(2010)

引用 0|浏览1
暂无评分
摘要
An implantable cardioverter defibrillator (ICD) is designed to sense life-threatening arrhythmias and terminate them, either by rapid pacing or by delivering an electrical shock. Nowadays it is a proven therapy for both primary and secondary prevention of sudden cardiac death. The typical configuration of an ICD consists of a right sensing/defibrillator lead with two coils (one distal, located in the right ventricle, and one proximal, located at the superior vena cava-right atrium junction) and an active can, the so-called ventricular triad. Although effective in the vast majority of patients, it could be argued that this is not the most rational arrangement in electrical terms, since the main shock vector is anteriorly displaced in relation to the greater portion of the left mass. We describe a case of an ICD defibrillation failure that was solved by placing an additional defibrillator lead in a tributary of the coronary sinus.