Effet de l’élution in vivo du technétium sur les volumes globulaires mesurés à l’aide d’hématies autologues marquées au 99m Tc : comparaison avec la technique au 51 Cr

Annales De Biologie Clinique(2020)

引用 2|浏览2
暂无评分
摘要
The purpose of this work was to compare the measured red-cell volume (RCV) using sodium pertechnetate [RCV-99mTc] compared to the reference technique using sodium radiochromate [RCV-51Cr] and to assess the influence of technetium-99 elution on the RCV-99mTc value. Ten patients had simultaneous measurements of RCV-99mTc and RCV-51Cr. Elution of Tc-99m from red blood cells was 2.9% and led to an average overestimation of RCV-99mTc of 3.7%. The introduction of individual tracer elution rates in the RCV-99mTc calculation corrects this overestimation.