Effects of iodinated contrast agent on HU-based dose calculation and dose delivered in iridium-192 high-dose-rate brachytherapy

JOURNAL OF CONTEMPORARY BRACHYTHERAPY(2022)

引用 0|浏览4
暂无评分
摘要
Purpose: This study compares the effect of iodinated contrast agent on Hounsfield unit (HU)-based TG-186 dose calculation vs. delivered dose for high-dose-rate (HDR) iridium-192 brachytherapy using a phantom model. Material and methods: A reservoir filled with a diluted contrast agent was placed inside a water phantom. A single steel needle applicator was centrally positioned inside the reservoir. Computed tomography (CT) datasets of five different contrast agent dilutions (25 to 300 mg/ml iodine concentration) were acquired, and dose calculations were performed with TG-186 ACE dose calculation formalism of Oncentra (R) Brachy (Elekta). The dose was measured with a PinPoint (R) ionization chamber (PTW) inside the contrast agent. ACE calculated and measured data were compared. Results: For the different contrast agent dilutions, averaged Hounsfield units from 453 +/- 21 to 2623 +/- 221 were obtained. Electron densities derived from CT data were significantly higher than corresponding electron densities calculated from chemical compositions. Consequently, the measured dose was higher than corresponding HU-based calculated dose. Relative deviation ranged from 2.5% to 7% per 10 mm penetration depth, depending on contrast agent concentration. Conclusions: The application of HU-based TG-186 dose formalisms in the presence of high-Z contrast agent bulks overestimates electron densities. Consequently, HU-based dose calculations result in a higher delivered dose than expected from the treatment planning system.
更多
查看译文
关键词
dose calculation, contrast agent, TG-186, Hounsfield unit, electron density
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要