Automatic exposure compensation of digital radiographic technologies does not affect alveolar bone-level measurement

Oral radiology(2022)

引用 0|浏览8
暂无评分
摘要
Objective To evaluate whether the automatic exposure compensation in the presence of high-density materials can affect the measurement of alveolar bone level. Methods Thirty regions of seven dry skulls and six mandibles were radiographed with and without a high-density material, using two digital radiographic technologies: photostimulable phosphor plate (PSP, Digora Optime) and sensor (CMOS, Digora Toto), totaling 120 images. The distances from the cement–enamel junction to the alveolar bone crest were measured using cone-beam computed tomography (CBCT) images to represent the reference standard. The same measurements of alveolar bone level and the average of the pixel values of the image were evaluated on the radiographs. Paired t test compared the average pixel values and alveolar bone-level measurements between images with and without high-density material. One-way analysis of variance compared the difference between radiographic and CBCT measurements ( α = 0.05). Results The high-density material reduced the pixel values in PSP ( p = 0.002) and CMOS ( p < 0.001) technologies, demonstrating the AEC functioning in both technologies. There was no difference in bone-level measurements between the images without and with the high-density material for both technologies ( p ≥ 0.091), or between the tomographic and radiographic measurements ( p ≥ 0.319). Conclusion In the presence of high-density material, the automatic exposure compensation reduces the average pixel values of the images (i.e., images get darker), but does not influence the radiographic measurements of alveolar bone level.
更多
查看译文
关键词
Alveolar bone level,Diagnostic imaging,Digital radiography,Radiographic image enhancement
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要