Radiation Dose Reduction in Paranasal Sinus CT

Otolaryngology–Head and Neck Surgery(2016)

引用 14|浏览25
暂无评分
摘要
Objectives To (1) compare the radiation dose of low-dose computed tomography (CT) to that of standard-dose CT, (2) determine the minimum optimal radiation dose for use in patients who need endoscopic sinus surgery, and (3) assess the reliability of iterative model reconstruction. Study Design Prospective single-institution study. Setting Tertiary care center. Subjects and Methods We recruited 48 adults with medically refractory sinusitis. Each patient underwent 4 scans with different CT parameters: 120 kV and 100 mAs (standard dose), 100 kV and 40 mAs (low dose), 100 kV and 20 mAs (very low dose), and 100 kV and 10 mAs (ultra-low dose). All CT scans were reconstructed via filtered back-projection, and ultra-low dose scans were additionally reconstructed through iterative model reconstruction. Radiation dose, image quality, and diagnostic performance were compared among the scans. Results Radiation doses decreased to 6% (ultra-low dose), 12% (very low dose), and 22% (low dose) of the standard-dose CT. The image quality of low-dose CT was similar to that of standard-dose CT. Ultra-low-dose CT with iterative model reconstruction was inferior to standard-dose CT for identifying anatomic structures, except for the optic nerve. All CT scans had 100% agreement for diagnosing rhinosinusitis. Conclusions With low-dose CT, the radiation dose can be decreased to 22% of that of standard-dose CT without affecting the image quality. Low-dose CT can be considered the minimum optimal radiation for patients who need surgery. Iterative model reconstruction is not useful for assessing the anatomic details of the paranasal sinus on CT.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要