Subcutaneous implantable cardioverter-defibrillators: long-term results of the EFFORTLESS study

EUROPEAN HEART JOURNAL(2022)

引用 39|浏览13
暂无评分
摘要
Aims To report 5-year outcomes of EFFORTLESS registry patients with early generation subcutaneous implantable cardioverter-defibrillator (S-ICD) devices. Methods and results Kaplan-Meier, trend and multivariable analyses were performed for mortality and late (years 2-5) complications, appropriate shock (AS) and inappropriate shock (IAS) rates. Nine hundred and eighty-four of 994 enrolled patients with diverse diagnoses (28% female, 48 +/- 17 years, body mass index 27 +/- 6 kg/m(2), ejection fraction 43 +/- 18%) underwent S-ICD implantation. Median follow-up was 5.1 years (interquartile range 4.7-5.5 years). All-cause mortality was 9.3% (95% confidence interval 7.2-11.3%) at 5 years; 703 patients remained in follow-up on study completion, 171 withdrew including 87 (8.8%) with device explanted, and 65 (6.6%) lost to follow-up. Of the explants, only 20 (2.0%) patients needed a transvenous device for pacing indications. First and final shock efficacy for discrete ventricular arrhythmias was consistent at 90% and 98%, respectively, with storm episode final shock efficacy at 95.2%. Time to therapy remained unaltered. Overall 1- and 5-year complication rates were 8.9% and 15.2%, respectively. Early complications did not predict later complications. There were no structural lead failures. Inappropriate shock rates at 1 and 5 years were 8.7% and 16.9%, respectively. Self-terminating inappropriately sensed episodes predicted late IAS. Predictors of late AS included self-terminating appropriately sensed episodes and earlier AS. Conclusion In this diverse S-ICD registry population, spontaneous shock efficacy was consistently high over 5 years. Very few patients underwent S-ICD replacement with a transvenous device for pacing indications. Treated and self-terminating arrhythmic episodes predict future shock events, which should encourage more personalized device optimization.
更多
查看译文
关键词
Implantable cardioverter-defibrillator, Subcutaneous ICD, Sudden death, Primary prevention, Secondary prevention
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要