Safety study of malapposition of the bio-corrodible nitrided iron stent in vivo

NANOTECHNOLOGY REVIEWS(2021)

引用 1|浏览10
暂无评分
摘要
To evaluate the safety of stent malapposition of corrodible nitride iron stent as biodegradable cardiovascular implants, a total of 108 stents were implanted into the abdominal aortas, iliac arteries, and iliac artery bifurcations of 36 New Zealand white rabbits separately. Each rabbit was implanted with three stents. After a follow-up period of 3 months, no thrombus and embolism were found in local and downstream vessels. And no other adverse events occurred either. Stent strut covered by endothelial layer started to show signs of degradation, while struts exposed to bifurcated blood flow covered by a layer of tissue and no rust particle was found on the surface. Also, there were no traces of thrombosis and traces of excess inflammation. The authors conclude that the risk brought by stent malapposition in less than 9 months is acceptable.
更多
查看译文
关键词
bio-absorb iron stent,corrosion,iron over-load,stent malapposition,thrombosis