Long-term durability of a new surgical aortic valve: A 1 billion cycle in vitro study

JTCVS Open(2022)

引用 7|浏览0
暂无评分
摘要
The Inspiris Resilia aortic valve demonstrated very good durability and hemodynamic performance after an equivalent of 25 years of simulated in vitro accelerated wear. The study valves exceeded 1 billion cycles of simulated wear, 5 times longer than the standard requirement for a tissue valve as stipulated in ISO 5840-2:2015.
更多
查看译文
关键词
valve durability,aortic valve,surgical heart valve,structural valve degeneration
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要