Single shot spiral TSE with annulated segmentation

MAGNETIC RESONANCE IN MEDICINE(2022)

引用 3|浏览6
暂无评分
摘要
Purpose To develop, optimize, and implement a single shot spiral turbo spin echo (TSE) sequence at 3T and to demonstrate its feasibility to acquire artifact free images of the central nervous system with 1 mm spatial resolution in Spiral TSE sequences with annulated spiral segmentation have been implemented with different acquisition modes. In fixed mode, the duration of each spiral segment is fixed to fill the available acquisition time t(acq). In tangential mode, the beginning of each spiral segment is reached via a straight tangential trajectory. Tangential mode allows faster transition and thus longer t(acq) for a given echo spacing (ESP), but less data points can be acquired per acquisition interval. Alternating between spiral-in and spiral-out readout in alternating echoes leads to a somewhat different point spread function for off-resonant spins. Results Images of the brain with 1 mm spatial resolution acquired with a variable density spiral with similar to 40% undersampling can be acquired in a single shot. All acquisition modes produce comparable image quality. Only mild artifacts in regions of strong susceptibility effects can be observed for ESP of 10 ms and below. The use of variable flip angle schemes allows seamless acquisition of consecutive slices and/or dynamic scans without waiting time between consecutive acquisitions. Comparison with images acquired at 1.5T shows reduced susceptibility artifacts within the brain and facial structures. Conclusion Single shot spiral TSE has been demonstrated to enable highly efficient acquisition of high-resolution images of the brain in <200 ms per slice.
更多
查看译文
关键词
FSE, pulseq, RARE, single shot MRI, spiral, spiral imaging, TSE, TSE
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要