Helical Needle Suture Inserter Provides Increased Suture Retention Force Compared To A Straight Needle
PROCEEDINGS OF THE 2021 DESIGN OF MEDICAL DEVICES CONFERENCE (DMD2021)(2021)
摘要
This paper presents an evaluation of the effect of needle geometry on the strength of a tether made using a barbed suture inserted into phantom tissue using a unique device. This tether is designed to secure an intrauterine device (IUD) to uterine fundus, with the aim of improving retention of IUDs inserted in the immediate postpartum period. A factorial experiment was designed to evaluate the effect of needle geometry on tether strength. Tether strength was characterized by the peak retention force of a suture subjected to a uniaxial tensile load. Experiments were performed using phantom tissue. Two needle geometries and three suture sizes were evaluated. Sutures deposited in phantom tissue with the helical needle had up to 132% increase in retention forces compared to sutures inserted with a straight needle, with more advantage at greater active length. The helical needle provides increased suture retention force and is a suitable tether delivery mechanism for this application.
更多查看译文
关键词
Needle, Barbed suture, Suture/tissue pull out test, Tether, Phantom tissue
AI 理解论文
溯源树
样例
![](https://originalfileserver.aminer.cn/sys/aminer/pubs/mrt_preview.jpeg)
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要