Ophthalmic Laser System Integrated With Speckle Variance Optical Coherence Tomography For Real-Time Temperature Monitoring

OPTICAL COHERENCE TOMOGRAPHY AND COHERENCE DOMAIN OPTICAL METHODS IN BIOMEDICINE XXII(2018)

引用 1|浏览2
暂无评分
摘要
A commercial ophthalmic laser system (R;GEN, Lutronic Corp) was integrated with a swept-source optical coherence tomography (OCT) imaging system for real-time tissue temperature monitoring. M-scan OCT images were acquired during laser-pulse radiation, and speckle variance OCT (svOCT) images were analyzed to deduce temporal signal variations related to tissue temperature change from laser-pulse radiation. A phantom study shows that svOCT magnitude increases abruptly after laser pulse radiation and recovered exponentially, and the peak intensity of svOCT image was linearly dependent on pulse laser energy until it saturates. A study using bovine iris also showed signal variation dependence on the laser pulse radiation, and the variation was more distinctive with higher energy level.
更多
查看译文
关键词
Optical coherence tomography, Ophthalmic laser system, temperature monitoring