Discordant Grading Of Aortic Stenosis Severity: New Insights From An In Vitro Study

STRUCTURAL HEART-THE JOURNAL OF THE HEART TEAM(2019)

引用 2|浏览4
暂无评分
摘要
Background : Low gradient (mean gradient, < 40 mmHg) aortic stenosis (AS) may occur in conjunction with a small valve effective orifice area (EOA) < 1 cm(2). This discordant grading in AS severity raises challenges from both a diagnostic and therapeutic standpoint. We sought to identify echocardiographic markers of severe AS in normal and low flow conditions in vitro.Methods : In an in vitro mock circulatory system, we tested 6 degrees of AS severity (3 severe and 3 nonsevere), and 3 levels of stroke volume (30, 50, 70 ml), heart rate (60, 80, 100 bpm), and mean arterial pressure (100, 120, 140 mmHg). MG and EOA were measured by Doppler-echocardiography.Results : In normal flow conditions, a MG of 40 mmHg corresponded to an EOA of 0.85 cm(2), and an EOA of 1.0 cm(2) to a MG of 28 mmHg. The presence of EOA < 0.85 cm(2) and MG between 24 and 40 mmHg in the low-flow condition yielded a 100% specificity for severe AS. In normal flow conditions, these thresholds were respectively < 0.9 cm(2) and > 36 mmHg.Conclusion : Low gradient AS may occur in severe AS in both low and normal flow conditions. The presence of a valve EOA < 0.85 cm(2) with an MG 25 mmHg may be used as a criteria to confirm severe AS in presence of low flow. In patients with normal-flow, a MG >= 37 mmHg is generally suggestive of severe AS.Abbreviations: AS: aortic stenosis; EOA: effective orifice area; HR: heart rate; LFLG: low-flow low-gradient; LVEF: left ventricular ejection fraction; MAP: mean arterial pressure; MG: mean gradient; NFLG: normal-flow low-gradient; SV: stroke volume
更多
查看译文
关键词
Aortic stenosis, in vitro, low flow, low gradient, echocardiography
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要