A 52 Mu W Heart-Rate Measurement Interface Fabricated On A Flexible Foil With A-Igzo Tfts

Mohammad Zulqarnaini,Stefano Stanzione, Jan-Laurens P. J. Van Der Steen,Gerwin H. Gelinck,Kris Myny,Sahel Abdinia,Eugenio Cantatore

ESSCIRC 2018 - IEEE 44TH EUROPEAN SOLID STATE CIRCUITS CONFERENCE (ESSCIRC)(2018)

引用 0|浏览0
暂无评分
摘要
This paper presents a 52 mu W heart-rate measurement interface for wearable applications, fabricated on a flexible foil with a-IGZO TFTs. The interface consists of a cascaded diode-connected load preamplifier with a gain of 22 dB in 3 kHz bandwidth. The output is provided as a binary PWM waveform, which can be sent to a reader device using wireless communication. Voltage to pulse-width conversion is achieved with a simple and power minimalistic approach, which exploits a reset integrator. Experimental characterization demonstrates an overall low-frequency gain of the system equal to 25 msec/V and an input referred noise of 186.3 mu V-rms. In-vivo measurements have been performed, resulting in a 60 bpm heart-rate for a person at rest.
更多
查看译文
关键词
IGZO,bio-signal read out,1/f noise,voltage to pulse width conversion,PWM