Gamma Ray Measurements At Omega With The Newest Gas Cherenkov Detector "Gcd-3"

A. M. Mcevoy,H. W. Herrmann,Y. Kim,A. B. Zylstra,C. S. Young,V. E. Fatherley,F. E. Lopez,J. A. Oertel, T. J. Sedillo, T. N. Archuleta, R. J. Aragonez,R. M. Malone, C. J. Horsfield,M. Rubery,S. Gales, A. Leatherland, W. Stoeffl,M. Gatu Johnson,W. T. Shmayda,S. H. Batha

9TH INTERNATIONAL CONFERENCE ON INERTIAL FUSION SCIENCES AND APPLICATIONS (IFSA 2015)(2016)

引用 13|浏览0
暂无评分
摘要
Initial results from the newest Gas Cherenkov Detector (GCD-3) are reported demonstrating improved performance over previous GCD iterations. Increased shielding and lengthening of the Cherenkov photon optical path have resulted in a diminished precursor signal with increased temporal separation between the precursor and the primary DT Cherenkov signal. Design changes resulted in a measured GCD-3 sensitivity comparable to GCD-1 at identical 100 psia CO2 operation. All metal gasket seals and pressure vessel certification to 400 psia operation allow for a GCD-3 lower Cherenkov threshold of 1.8 MeV using the fluorinated gas C2F6 as compared to the 6.3 MeV lower limit of GCD-1 and GCD-2. Calibration data will be used to benchmark GEANT4 and ACCEPT detector models. The GCD-3 acts as a prototype for the Super GCD being fielded at the National Ignition Facility (NIF) as part of the National Diagnostics Plan and will be installed at NIF in early 2016.