Investigation Of The Neutron-Gamma Ray Discrimination Performance At Low Neutron Energy Of A Solution-Grown Stilbene Scintillator

ADVANCEMENTS IN NUCLEAR INSTRUMENTATION MEASUREMENT METHODS AND THEIR APPLICATIONS (ANIMMA 2019)(2020)

引用 2|浏览2
暂无评分
摘要
The results of experiments performed with a O25 x 25 mm solution-grown stilbene crystal in mono-energetic neutron fields in the 80-to-230 keV energy range are presented. The goal of the measurements, performed at the AMANDE facility, was to explore the capabilities of this organic scintillator to measure neutrons at the lowest possible energy with good pulse shape discrimination (PSD). The time of flight (TOF) technique was used in order to help with the neutron-gamma discrimination. The data are collected via a programmable digital acquisition (DAQ) system CAEN DT7530 using the software CoMPASS with the charge comparison method (CCM). The data are analysed using post-processing codes developed in the ROOT environment. The results show that the stilbene detector has discrimination capabilities for energies as low as 80 keV.
更多
查看译文
关键词
PSD, Organic scintillators, Solution-grown stilbene, Mono-energetic neutrons, Time of flight, digital acquisition system, AMANDE facility
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper