Bead-Based Digital Microfluidic Immunoassay For Il-1 Beta Detection In Embryo Culture Medium

2017 IEEE 12TH INTERNATIONAL CONFERENCE ON NANO/MICRO ENGINEERED AND MOLECULAR SYSTEMS (NEMS)(2017)

引用 0|浏览0
暂无评分
摘要
One of the challenges in IVF treatment is to improve the pregnancy rate of selective single embryo implantation. The concentration of IL-1 beta in embryo culture medium may help to predict pregnancy rate. Clinically, each embryo can only get small amount of sample size. However, the traditional immunoassay cannot detect biomarkers in such small sample sizes. Therefore, we combined immunoassay with the bead-based digital microfluidic chip. The chip has two advantages, low sample requirement and low detection limit. We established the mouse IL-1 beta calibration curve, confirm the Human Tubal Fluid (HTF) does not influence the test result and demonstrate that mouse IL-1 beta in clinical mouse embryo culture medium is positively correlate to cultured days. This device has a great potential for increasing IVF pregnancy rates.
更多
查看译文
关键词
digital microfluidic, immunoassay, embryo growth factor
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要