Inappropriate rate response in a leadless pacemaker due to automatic rate profile optimization
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY(2022)
摘要
A three-axis accelerometer sensor is incorporated in a leadless pacemaker (Micra, Medtronic(Inc)) for rate adaptation. Three sensor setpoints at lower rate (LR), activity of daily living rate (ADLR), and upper rate (UR) are used to convert detected acceleration to a desired rate response. We describe inappropriate rate acceleration at rest and during chest physiotherapy in a sedentary elderly patient with leadless pacemaker. The underlying mechanism and various programming options are discussed.
更多查看译文
关键词
atrial fibrillation, leadless pacemaker, tachycardia