In vivo human retinal swept source optical coherence tomography and angiography at 830 nm with a CMOS compatible photonic integrated circuit

SCIENTIFIC REPORTS(2021)

引用 3|浏览5
暂无评分
摘要
Photonic integrated circuits (PIC) provide promising functionalities to significantly reduce the size and costs of optical coherence tomography (OCT) systems. This paper presents an imaging platform operating at a center wavelength of 830 nm for ophthalmic application using PIC-based swept source OCT. An on-chip Mach–Zehnder interferometer (MZI) configuration, which comprises an input power splitter, polarization beam splitters in the sample and the reference arm, and a 50/50 coupler for signal interference represents the core element of the system with a footprint of only (12 × 5) mm^2 . The system achieves 94 dB imaging sensitivity with 750 W on the sample, 50 kHz imaging speed and 5.5 m axial resolution (in soft tissue). With this setup, in vivo human retinal imaging of healthy subjects was performed producing B-scans, three-dimensional renderings as well as OCT angiography. These promising results are significant prerequisites for further integration of optical and electronic building blocks on a single swept source-OCT PIC.
更多
查看译文
关键词
Integrated optics,Silicon photonics,Three-dimensional imaging,Science,Humanities and Social Sciences,multidisciplinary