Spectrum Medical Quantum or Terumo CDI 500: Which Device Measures Hemoglobin and Oxygen Saturation Most Accurately When Compared to a Benchtop Blood Analyzer?

The journal of extra-corporeal technology(2021)

引用 1|浏览1
暂无评分
摘要
To examine the accuracy between analyzers, the Terumo CDI 500 and the Spectrum Medical Quantum were compared to each other and to the ABL90 FLEX benchtop blood analyzer. Patients were retrospectively identified who underwent cardiac surgery requiring cardiopulmonary bypass between August 1, 2018 and November 1, 2019. Hemoglobin and venous saturation (SvO2) values from all three analyzers were collected. Measurements from the Quantum and the CDI 500 were averaged over 1 minute to provide a single value for the minute for the given device. Blood analysis on the ABL90 benchtop device was performed at a minimum of every hour during congenital cardiopulmonary bypass (CPB). There were 519 patients included in the analysis. Data points numbering 69,404 and 70,598 were analyzed when comparing the CDI 500 to the Quantum for hemoglobin and SvO2, respectively. Comparison of hemoglobin and SvO2 for the CDI 500 and Quantum versus ABL90 used 2283 and 1414 data points respectively, in each group. The CDI 500 and Quantum reported hemoglobin within 1 g/dL of the ABL90 86.9% and 87.5% of the time, respectively. The CDI 500 and Quantum reported SvO2 within 3% of the ABL90 61.0% and 57.9% of the time, respectively. The mean difference between the CDI 500 and Quantum hemoglobin and SvO2 measurements equaled .194 g/dL (p < .001) and .861% (p < .001), respectively and were both significantly different from zero. All device comparisons were statistically significantly different when compared to zero difference, likely due to the large data set as the magnitudes of these differences are all quite small and may not be clinically significant. However, while the reader should judge for themselves based upon their specific practice, in our opinion, the 95% Limit of Agreement was too large for either the CDI 500 or Quantum hemoglobin and SvO2 values to be substituted for ABL90 values. As recommended by the manufacturers, the CDI 500 and Quantum should only be used as a trending device.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要