Clinical monitoring of activated clotting time during cardiothoracic surgery: comparing the Hemochron ® Response and Hemochron ® Signature Elite.

Perfusion(2023)

引用 2|浏览0
暂无评分
摘要
The Elite was more reliable, but the variation was higher for the Elite than the Response. The observed positive bias in the Response compared to the Elite could affect heparin administration during surgery making the two systems not interchangeable.
更多
查看译文
关键词
activated clotting time,bias,cardiopulmonary bypass,cardiothoracic surgery,heparin,variability