Variation In Dosimeter Calibration Factor (Ndw) Over A Period Of 20 Years

JOURNAL OF MEDICAL PHYSICS(2021)

引用 0|浏览3
暂无评分
摘要
Long delays in renewal of calibration of secondary standards radiation dosimeters in radiation oncology centers due to the COVID19 pandemic have aroused concerns regarding accuracy in dose delivery to patients. The concerns are mainly due to the uncertainty in the absorbed dose to water calibration factor (N-DW) over a period of time. In this study, the N(DW)factor for two ion chambers, thimble type (Farmer) and parallel plate type (Markus), used in most of the radiotherapy centers, were retrospectively reviewed for 20 years. The calibration on all occasions except once was carried out at the Secondary Standards Dosimetry Laboratory, Bhabha Atomic Research Centre, Mumbai. The change in the N(DW)factor over this period was < +/- 3%. We, therefore, believe that a dosimeter with no history of repairs showed reasonable stability in the N(DW)factor over a long period.
更多
查看译文
关键词
Calibration factors, farmer chamber, Markus chamber, secondary standards dosimeter
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要