Design, calibration, and application of a cryogenic low-background infrared radiometer for spectral irradiance and radiance measurements from 4 μm to 20 μm wavelength.

Optical engineering (Redondo Beach, Calif.)(2021)

引用 0|浏览0
暂无评分
摘要
We have constructed, calibrated, and tested a cryogenic low-background infrared radiometer for both spectral radiance and irradiance measurements over the 4 μm to 20 μm wavelength range. The primary purpose of the Missile Defense Transfer Radiometer (MDXR) is to measure absolute irradiance or radiance from cryogenic infrared test chamber sources using a photoconductive Si:As Blocked Impurity Band (BIB) detector and a set of spectral filters. The MDXR also includes an absolute cryogenic radiometer (ACR) and a Fourier transform spectrometer (FTS). For irradiance measurements, the ACR is used to provide the primary power scale for the BIB detector in conjunction with spectral filters, while the FTS/BIB configuration derives its scale from an internal blackbody source. The two measurement scales show agreement for the irradiance of highly collimated (< 1 mrad) infrared beams from 10-13 W/μm/cm2 to 10-8 W/μm/cm2 within the combined relative uncertainties of 2.6 % (coverage factor k = 1.) We have also calibrated the radiometer for radiance measurements by using a large cavity fluid bath blackbody that overfills the spatial and angular extent of the radiometer entrance pupil. The radiometric calibration uncertainty analysis of the radiometer as well as its maintenance and stability are discussed.