Micra-Av Leadless Pacemaker And Atrioventricular (Dys)Synchrony: A Stepwise Process

PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY(2021)

引用 8|浏览2
暂无评分
摘要
Leadless pacemakers have become a major breakthrough in the management of bradyarrhythmia as an attractive alternative to the standard transvenous pacemakers, but safety and long-term outcomes of the new Micra-AV, which is capable guarantee atrioventricular (AV) synchrony, have been poorly described. Hence, we describe how we managed a 57-year-old patient who developed pacemaker syndrome due to significant Micra-AV dyssynchrony. This case emphasizes how leadless pacemaker AV synchrony could be overestimated, thereby requiring a stepwise process leading to adequate device reprogramming. Holter-ECG monitoring and exercise test resulted to be valuable tools for an early detection of inadequate AV synchrony, integrating and completing device reports.
更多
查看译文
关键词
atrioventricular synchronous leadless pacemaker, atrioventricular synchrony, leadless pacemaker, micra-AV