Functional Properties Of Polyurethane Ureteral Stents With Plga And Papaverine Hydrochloride Coating

INTERNATIONAL JOURNAL OF MOLECULAR SCIENCES(2021)

引用 8|浏览19
暂无评分
摘要
Despite the obvious benefits of using ureteral stents to drain the ureters, there is also a risk of complications from 80-90%. The presence of a foreign body in the human body causes disturbances in its proper functioning. It can lead to biofilm formation on the stent surface, which may favor the development of urinary tract infections or the formation of encrustation, as well as stent fragmentation, complicating its subsequent removal. In this work, the effect of the polymeric coating containing the active substance-papaverine hydrochloride on the functional properties of ureteral stents significant for clinical practice were assessed. Methods: The most commonly clinically used polyurethane ureteral Double-J stent was selected for the study. Using the dip-coating method, the surface of the stent was coated with a poly(D,L-lactide-glycolide) (PLGA) coating containing the papaverine hydrochloride (PAP). In particular, strength properties, retention strength of the stent ends, dynamic frictional force, and the fluoroscopic visibility of the stent during X-ray imaging were determined. Results: The analysis of the test results indicates the usefulness of a biodegradable polymer coating containing the active substance for the modification of the surface of polyurethane ureteral stents. The stents coated with PLGA+PAP coating compared to polyurethane stents are characterized by more favorable strength properties, the smaller value of the dynamic frictional force, without reducing the fluoroscopic visibility.
更多
查看译文
关键词
ureteral stent, PLGA with papaverine hydrochloride coatings, retention strength, break strength, dynamic frictional force, radiopacity
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要