Evaluation Of Identification Accuracy Using Aiqs For Gc-Ms For Measuring Heavily Contaminated Samples

CHEMOSPHERE(2021)

引用 3|浏览8
暂无评分
摘要
Automated identification and quantification systems with gas chromatography-mass spectrometry (GC-MS) (i.e., AIQS-GC) are used as a simple and comprehensive method for screening chemicals existing in the environment and are expected to be useful for emergency surveys in the event of a disaster. However, reports on the potential of AIQS-GC in heavily contaminated samples (HCSs) are limited. In this study, the identification performance of AIQS-GC was confirmed by comparing the exact mass of the targets identified by AIQS-GC with the measured accurate mass using GC-quadrupole-time-of-flight MS (GC-QTofMS) and by employing firefighting wastewater as HCS. In HCS, the mass spectrum interference was determined to cause false positives. The GC-QTofMS method revealed the presence of false positives and the false rate of AIQS-GC in HCS. Herein, AIQS-GC showed high identification accuracy in a normal sample such as river water. Conversely, in HCS, AIQS-GC may lead to incorrect evaluations. The combination of AIQS-GC and support method using GC-QTofMS, which can avoid the false positive is extremely useful for the rapid and easy analysis of HCS.
更多
查看译文
关键词
Heavily contaminated sample, Automated identification and quantification, system (AIQS), Identified accuracy, GC-QTofMS, Mass error, Comprehensive screening analysis