New Polarized Neutron Diffraction Setup for Precise High-Field Investigations of Magnetic Structures up to 8 T at MLZ

IEEE TRANSACTIONS ON MAGNETICS(2022)

引用 0|浏览15
暂无评分
摘要
New actively shielded asymmetric vertical field split-coil superconducting magnet with maximal field of 8 T has been recently procured and implemented on the beam line polarization investigator (POLI) at Maier-Leibnitz Zentrum (MLZ). The magnet is designed in order to facilitate the single crystal diffraction with polarized neutrons. In order to provide high neutron polarization in the vicinity of the new strong magnet, a dedicated compact-size solid-state supermirror bender polarizer optimized for the short neutron wavelength on POLI has been built. A Mezei flipper, which had been previously developed for lower field, could be upgraded with additional guide fields in order to be used in combination with new polarizer and magnet. All components are discussed and the performance test for the entire setup is presented. High efficiency of the new polarized neutron diffraction (PND) setup on POLI is demonstrated. PND experiments in the fields up to 8 T are now available at MLZ.
更多
查看译文
关键词
Neutrons, Superconducting magnets, Magnetic shielding, Magnetic noise, Diffraction, Magnetic domains, Perpendicular magnetic anisotropy, High magnetic field, magnetic structure, polarized neutron diffraction (PND), supermirror polarizer
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要