Use of a fluoropolymer-based paclitaxel-eluting stent for arteriovenous graft outflow vein stenosis in hemodialysis patients

Journal of Vascular Surgery Cases, Innovations and Techniques(2021)

引用 1|浏览1
暂无评分
摘要
We implanted a fluoropolymer-based paclitaxel-eluting stent (FP-PES) in four hemodialysis patients with refractory outflow venous stenosis of their arteriovenous graft. The mean observation period after FP-PES implantation was 11.5 ± 4.7 months (range, 7.0-18.0 months). After FP-PES implantation, the patients were evaluated by ultrasound every 3 months. No of the patients experienced neointimal hyperplasia in the stents during the observation period, and no reintervention was performed. FP-PESs could be an attractive alternative to percutaneous transluminal angioplasty for patients with refractory outflow venous stenosis of arteriovenous hemodialysis grafts.
更多
查看译文
关键词
Arteriovenous shunt,Drug-eluting stents,Hemodialysis,Paclitaxel,Stenosis