2-Mercaptoethanol (2-Me)-Based Iats Or Polybrene Method Mitigates The Interference Of Daratumumab On Blood Compatibility Tests

HEMATOLOGY(2021)

引用 4|浏览7
暂无评分
摘要
ObjectivesTreating red blood cells (RBCs) with dithiothreitol (DTT) is a wildly-recommended to overcome the interference of the daratumumab (DARA) with blood compatibility testing. Nevertheless, DTT can be hard to obtain in the clinical laboratory, while its use in routine practice may be time-consuming. In the following study, we explored the feasibility of using a commercial 2-mercaptoethanol (2-ME) working solution or the time-saving Polybrene method to mitigate DARA interference.MethodsAntibody screening and cross-matching were performed using 2-ME or DTT-based indirect antiglobulin tests (IATs) and Polybrene method (with human IgG anti-E same IATs titer as DARA as positive control) on 37 samples. Most clinically important blood group antigens on RBCs were detected after treatment with 2-ME or DTT.ResultsTreating RBCs with 2-ME eliminates the DARA interference with the antibody screening or cross-matching; yet, K antigen is denatured during treatment. DARA does not interfere with antibody screening and cross-matching via Polybrene method, while 2+ agglutinations of anti-E antibody with the same titer (IATs method) as DARA could be observed in the positive controls via this method.Conclusion2-ME-based IATs or Polybrene method could replace DTT-based IATs to mitigate DARA interference.
更多
查看译文
关键词
Daratumumab, anti-CD38, 2-mercaptoethanol, polybrene, blood transfusion, dithiothreitol, blood compatibility testing, indirect antiglobulin test