Method Optimization And Analysis Of Flurbiprofen Loaded Eudragit L100 Nanoparticles Using Rp-Hplc Technique: A Central Composite Design Approach

Materials Today: Proceedings(2021)

引用 8|浏览0
暂无评分
摘要
For the development and analysis of Flurbiprofen (FBP) loaded Eudragit L100 nanoparticles the Central composite design (CCD), a response surface methodology, was used for optimization the HPLC process. The desired method development was made on the HPLC gradient System with auto injector. A stationary phase, composed of C18 (AGILENT) with a sample size of 20ll and a particle size of 5ll, was included in the method. The column size was 4.6 x 250 mm with temperature 22.5 degrees C at 247 nm was detected by the analytes. Acetonitrile (ACN): Methanol (MeOH): 0.05% ortho phosphoric acid in water (82.5: 5: 12.5) with a pH value of 3. The linearity, precision, accuracy, repeatability, recovery, robustness, and ANOVAs test, the approach mentioned was validated with good results and coefficient correlation (R2) 0.994. Its acceptance conditions were met with the outcome of each parameter.(c) 2021 Elsevier Ltd. All rights reserved.Second International Conference on Aspects of Materials Science and Engineering (ICAMSE 2021).
更多
查看译文
关键词
Flurbiprofen, Central composite design, Nanoparticles, RP-HPLC, Method development, Validation