Dosimetry Of A 6 Mv Flattening Filter-Free Small Photon Beam Using Various Detectors

BIOMEDICAL PHYSICS & ENGINEERING EXPRESS(2021)

引用 2|浏览1
暂无评分
摘要
The present study aimed to dosimetrically evaluate the small-fields of a 6 MV flattening filter-free (FFF) photon beam using different detectors.The 6 MV FFF photon beam was used for measurement of output factor, depth dose, and beam profile of small-fields of sizes 0.6 cm x 0.6 cm to 6.0 cm x 6.0 cm. The five detectors used were SNC125c, PinPoint, EDGE, EBT3, and TLD-100. All measurements were performed as per the International Atomic Energy Agency TRS 483 protocol. Output factors measured using different detectors as direct reading ratios showed significant variation for the smallest fields, whereas after correcting them according to TRS 483, all sets of output factors were nearly compatible with each other when measurement uncertainty was also considered. The beam profile measured using SNC125c showed the largest penumbra for all field sizes, whereas the smallest was recorded with EDGE. Compared with that of EBT3, the surface dose was found to be much higher for all the other detectors. PinPoint, EBT3, TLD-100, and EDGE were found to be the detector of choice for small-field output factor measurements; however, PinPoint needs special attention when used for the smallest field size (0.6 cm x 0.6 cm). EDGE and EBT3 are optimal for measuring beam profiles. EBT3, PinPoint, and EDGE can be selected for depth dose measurements, and EBT3 is suitable for surface dose estimation.
更多
查看译文
关键词
small-field dosimetry, FFF photon beam, detector, output factor