Bilateral Comparison Of 1 Omega And 10 K Omega Standards (Ongoing Bipm Key Comparisons Bipm.Em-K13.A And 13.B) Between The Nmc A*Star (Singapore) And The Bipm

B Rolland,M Stock, S W Chua, K H Lim, E Tan

METROLOGIA(2020)

引用 0|浏览1
暂无评分
摘要
This report gives the result of a bilateral comparison of resistance between the NMC A*STAR (Singapore) and the BIPM carried out in 2018-2019. Two 1 Ω and two 10 kΩ travelling standards belonging to the BIPM were used. The comparison was carried out with an 'A-B-A' pattern of measurements; the standards were measured first at the BIPM for a period of about two months, then for a period of about one month at the NMC A*STAR, and finally again at the BIPM. The measurand was the 4 terminal dc resistance at low power. The BIPM was the pilot laboratory, and the comparison forms part of the ongoing BIPM key comparisons BIPM.EM-K13.a (for 1 Ω) and BIPM.EM-K13b (for 10 kΩ). The results from the NMC A*STAR and the BIPM were found to be in good agreement, with a difference smaller than the relative expanded uncertainty (95% confidence, k = 2) of 0.064 × 10−6 for 1 Ω and 0.036 × 10−6 for 10 kΩ. Main text To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/. The final report has been peer-reviewed and approved for publication by the CCEM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).