Rescue Of Failed Xen-45 Gel Implant By Nd:Yag Shock Wave To Anterior Chamber Tip To Dislodge Hidden Intraluminal Occlusion

JOURNAL OF GLAUCOMA(2021)

引用 4|浏览1
暂无评分
摘要
Purpose: The purpose of this study was to inform ophthalmic surgeons in a timely manner of the hidden problem of clear intraluminal cellular debris as a cause for XEN-45 failure and to describe low energy neodymium-doped yttrium aluminum garnet (Nd:YAG) laser revision with periluminal anterior chamber tip shockwave treatment to improve flow to the bleb. Patients and Methods: Six patients with visibly patent stent lumen post XEN-45 surgery. These eyes developed rising intraocular pressure (IOP) with a history of excellent prior bleb formation and were treated successfully with Nd:YAG laser shockwave therapy to disperse assumed intraluminal cellular debris. The laser was aimed just anterior and axial to the intracameral tip of the gel stent through a gonioscopy lens. Results: Six patients with an average age of 75 years (60 to 90 y), preoperative IOP of 30 mm Hg (16 to 52 mm Hg) on an average of 2 antiglaucoma medications (0 to 4) underwent periluminal anterior chamber tip shock wave at an average of 12 months (1 to 38 mo) from XEN-45 surgery. The IOP was immediately reduced to an average of 15 mm Hg (8 to 23 mm Hg) and last IOP averaged 15 mm Hg (10 to 23 mm Hg) on 1.5 medications (0 to 4) at 4 months post periluminal anterior chamber tip shock wave. Conclusion: Nd:YAG laser revision of hidden blockage of a XEN-45 gel implant with periluminal anterior chamber tip shockwave treatment can disperse invisible intraluminal cellular debris and improve flow in a failing XEN-45 microstent, especially when distal fibrosis is not excessive.
更多
查看译文
关键词
XEN-45 gel stent, intraluminal blockage, occult blockage of XEN-45 microstent, Nd, YAG laser revision, postoperative complication
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要