Towards high-rate RPC-based thermal neutron detectors using low-resistivity electrodes

JOURNAL OF INSTRUMENTATION(2021)

引用 2|浏览7
暂无评分
摘要
We present experimental results on the counting ratemeasurements for several single-gap B-10 lined resistive plate chambers (B-10-RPCs) with anodes made from standard float glass, low-resistivity glass and ceramic. The measurements were performed at the V17 monochromatic neutron beamline (3.35 angstrom) at the Helmholtz-Zentrum Berlin. For the B-10-RPCs with 0.28mm thick float glass a maximum counting rate density of about 8 x 10(3) Hz/cm(2) was obtained. In the case of low resistivity glass and ceramic, the counting rate density did not deviate from linear dependence on the neutron flux up to the maximum flux available at this beamline and exceeded a value of 3 x 10(4) Hz/cm(2).
更多
查看译文
关键词
Neutron detectors (cold, thermal, fast neutrons), Resistive-plate chambers, Instrumentation for neutron sources