An analytical model for electronic noise in a cryogenic bolometer detector readout circuit

V. Vatsa,A. Reza,A. Mazumdar, M. S. Pose, S. Mallikarjunachary,V. Nanal,R. G. Pillay,S. Ramakrishnan,A. Shrivastava

arxiv(2022)

引用 0|浏览8
暂无评分
摘要
This paper presents an analytical model to quantify the measured noise in a cryogenic bolometer readout circuit. The model includes the contributions from the bias resistors and sensor resistors, voltage and current noise of amplifier, and cable capacitance. The model parameters are empirically estimated using frequency domain analysis of the measured noise data of indigenously developed Neutron Transmutation Doped (NTD) Ge sensors. The model is shown to describe noise data for NTD Ge sensors over a wide range of resistances corresponding to temperatures in the range 20 - 70 mK. Relative contributions of different components are discussed and it is shown that the contribution to the overall noise from the differential amplifier at 300 K is the dominant source. It is observed that the amplifier flicker noise is significantly lower than that specified in the amplifier datasheet. The present study also indicates that a desirable value of resistance of NTD sensor (R$_{NTD}$) from noise considerations is $\leq$ 1 G$\Omega$ at $sim$ 20 mK.
更多
查看译文
关键词
Cryogenic detectors,Front-end electronics for detector readout,Double-beta decay detectors