Remote monitoring of vital signs in infectious patients minimizes the risks of viral transmissions to healthcare professionals. Donning face masks could reduce the risk of viral transmissions and is currently practiced in medical facilities. An acoustic-sensing device was attached to face masks to assist medical facilities in remotely monitoring patients' respiration rate and wheeze occurrence. Usability and functionality studies of the modified face mask were evaluated on 16 healthy participants. Participants were blindfolded throughout the data collection process. Respiratory rates of the participants were evaluated for one minute. The wheeze detection algorithm was assessed by playing 176 wheezes and 176 normal breaths through a foam mannequin. No discomfort was reported from the participants who used the modified mask. The mean error of respiratory rate was found to be 2.0 +/- 1.3 breath per minute. The overall accuracy of the wheeze detection algorithm was 91.9%. The microphone sensor that was first designed to be chest-worn has been proven versatile to be adopted as a mask attachment. The current findings support and suggest the use of the proposed mask attachment in medical facilities. This application can be especially helpful in managing a sudden influx of patients in the face of a pandemic.