Micro-computed tomography evaluation of general trends in aligner thickness and gap width after thermoforming procedures involving six commercial clear aligners: An in vitro study

KOREAN JOURNAL OF ORTHODONTICS(2021)

引用 13|浏览4
暂无评分
摘要
Objective: To assess the effects of thermoforming on aligner thickness and gap width in six aligner systems with the same nominal thickness. Methods: Six passive upper aligners of different brands were adapted to a single printed cast. Each sample was evaluated with high-resolution micro-computed tomography. To investigate aligner thickness and gap width, two-dimensional (2D) analysis was conducted assessing the effects of the following variables: tooth type (central incisor, canine, and first molar), 2D reference points, and aligner type. Data were analyzed and compared using analysis of variance and Tukey's post-hoc tests (p < 0.05). Results: Tooth type, dental region, and aligner type affected both the gap width and aligner thickness. The aligner thickness remained moderately stable across the arch only in the F22. Conclusions: All thermoformed samples displayed smaller aligner thickness and gap width at anterior teeth and both gingival and coronal centers than at posterior teeth and occlusal surfaces.
更多
查看译文
关键词
Gap width, Aligner thickness, Thermoforming procedure, Clear aligner therapy