Characterization of an electrically cooled BEGe detector till Eγ∼7 MeV

Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment(2020)

引用 2|浏览4
暂无评分
摘要
An electrically cooled Broad Energy Germanium (BEGe) detector has been characterized in the energy range Eγ∼0.122 - 7 MeV by utilizing the γ-rays emitted by a short-lived resonance state in 15O populated through 14N(p, γ) reaction and standard radioactive source (152Eu). The experimental results have been reproduced through simulations with GEANT4 code, including vendor specified detector geometry along with the detailed construction of the target holder flange, to delineate the effects of the holder at various energies and detector position. Later the efficiency with a bare point source has been simulated. It has been found that the electrically cooled BEGe detector is suitable for usage in the γ-ray spectroscopy as well as for the study of resonance phenomena in nuclear astrophysics.
更多
查看译文
关键词
Gamma detectors (BEGe, HPGe),Electrically cooled detector,Resonance reaction,Efficiency,Simulation of detector response,High energy gammas