Evaluations On The Stability And Bio-Compatibility Of A New Piezoelectric Microphone For The Implantable Middle Ear Microphone

ACTA OTO-LARYNGOLOGICA(2021)

引用 1|浏览3
暂无评分
摘要
BackgroundA new floating piezoelectric microphone (NFPM), driven by the acoustic vibration of the ossicles, has been manufactured in the lab.ObjectivesThis study aimed at exploring the stability and bio-compatibility of this NFPM.Materials and methodsThe NFPM was implanted into the cat tympanic cavity by clamping it to the handle of the malleus, and then a temporal bone computed tomography (CT) plain scan and three dimensional reconstruction were performed to observe its coupling with the handle of the malleus. After six months of implantation, cats' blood samples were taken for pro-inflammatory factor analysis. Then, the operative cavity was opened to check the NFPM, the auditory ossicular chain and some granulation tissue surrounding the NFPM.ResultsThe NFPM was firmly clamped to the handle of the malleus of the cats during the six months of implantation. Besides, there was no obvious systemic inflammatory response in the experimental animals. In addition, local proliferation of granulation tissue occurred in the tympanic cavity without hampering the movement of the auditory ossicle,or causing ischemia of the auditory ossicle.ConclusionThe NFPM could be implanted in our experimental cats for a long period of time and had good bio-compatibility.
更多
查看译文
关键词
Totally implantable cochlear implants, piezoelectric microphone, bio-compatibility