Entrapment Of Diagnostic Catheter Within Advisor Hd Grid Mapping Catheter

JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY(2021)

引用 5|浏览0
暂无评分
摘要
The potential for catheter entanglement with the HD Grid mapping catheter is explicitly stated in the manufacturer's product manual. A case of an entrapped 6 French quadripolar diagnostic catheter within an HD Grid mapping catheter is presented. We discuss the diagnosis, management, and resolution of this complication in our patient. The patient's arrhythmia was successfully eliminated, and no vascular complication in the postprocedural setting nor arrhythmia recurrence at follow-up were observed. Strategies to prevent and safely manage this complication, while salvaging access, are also discussed.
更多
查看译文
关键词
ablation, catheter entanglement, electrophysiology study, procedural complication, supraventricular tachycardia
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要