A Transfer Function Measurement Setup With an Electro-Optic Sensor for MR Safety Assessment in Cascaded Media

IEEE Transactions on Electromagnetic Compatibility(2021)

引用 5|浏览1
暂无评分
摘要
This article introduces a transfer function (TF) measurement setup for cascaded media (CM) and demonstrates that the full immersion of an implant in a lossy medium might result in an underestimation of the RF-induced heating compared to partial immersion. A CMTF measurement setup was constructed using medium-specific local electric field sources and electro-optical sensor (EOS). TFs of bare and in...
更多
查看译文
关键词
Wires,Media,Heating systems,Transmission line measurements,Lead,Probes,Earth Observing System