Determination of the mechanical properties of polymeric microneedles by micromanipulation

Journal of the Mechanical Behavior of Biomedical Materials(2021)

引用 20|浏览8
暂无评分
摘要
Precise characterization of the mechanical properties of polymeric microneedles is crucial for their successful penetration into skin and delivery of the loaded active ingredients. However, most available strategies for this purpose are based on compression of the whole patch, which only provide the average rupture force of the needles and can not give information on the variations across individual microneedles in the patch. In this study, we determined the mechanical strength of individual microneedles of two types of hyaluronic acid microneedles with or without loaded model drugs using a micromanipulation technique. The applied force as a function of displacement of the microneedles was recorded, which was used to determine the rupture displacement, rupture force, and then to derive and calculate normal stress-deformation curve, rupture stress and Young′s modulus of individual microneedles. The obtained data suggest that the molecular weight of the polymer and the loading of drug into the microneedles can significantly affect the rupture behavior and mechanical properties of the microneedles, which provides a foundation for preparing sufficiently strong microneedles for controlled drug delivery.
更多
查看译文
关键词
Mechanical properties,Polymeric microneedles,Micromanipulation,Rupture force,Normal stress