Position-Dependent Dysfunction Of Deep Brain Stimulation In Tourette Syndrome: Diagnostic Clues

TREMOR AND OTHER HYPERKINETIC MOVEMENTS(2019)

引用 1|浏览9
暂无评分
摘要
Background: Detection of defective deep brain stimulation (DBS) contacts/electrodes is sometimes challenging.Case Report: We report a patient with Tourette syndrome (TS), who presented with abrupt tic increase and mild generalized headache 9 years after DBS implantation. On the suspicion of a hardware defect, a fracture of the DBS electrode and extension lead was ruled out by radiography and standard implantable pulse generator readouts. Further investigation revealed position-dependent modifiable therapeutic impedances, suggesting an impaired contact of the extension lead/ adaptor. After replacement normal impedances were recorded, and the patient fully recovered.Discussion: In DBS dysfunction with inconspicuous hardware check, position-dependent defects might be suspected.
更多
查看译文
关键词
Deep brain stimulation, hardware complications, headache, impedance, Tourette syndrome