Improved Sensitivity Roll-Off In Dual Reference, Buffered Spectral-Domain Optical Coherence Tomography

JOURNAL OF BIOMEDICAL OPTICS(2021)

引用 1|浏览2
暂无评分
摘要
Significance: While spectral-domain optical coherence tomography (SD-OCT) is a preferred form of OCT imaging, sensitivity roll-off limits its applicability for certain biomedical imaging applications.Aim: The aim of this work is to extend the imaging range of conventional SD-OCT systems for imaging large luminal organs such as the gastrointestinal tract.Approach: We present an SD-OCT system operating at a center wavelength of 1300 nm that uses two delayed reference arms to reduce sensitivity roll-off and an optical switch and a fiber optic delay line to ensure that the interference spectra are acquired from the same sample time window.Result: The proposed system was used to image swine colon ex vivo and duodenum in vivo, demonstrating improved image quality due to a similar to 14 dB increase in sensitivity at the edges of the ranging depth.Conclusion: The proposed system requires modest hardware implementation and is compatible with catheter-based endoscopic helical scanning with enhanced sensitivity for the samples at a distance of similar to 6 mm from the zero delay point. (C) The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License.
更多
查看译文
关键词
optical coherence tomography, in vivo, gastrointestinal tract, tethered capsule, endomicroscopy