Standardization of 223Ra by live-time anticoincidence counting and gamma-ray emission determination.

Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine(2021)

引用 4|浏览3
暂无评分
摘要
From a commercial supplier, three independent lots of an aqueous solution containing 223Ra in equilibrium with its deteriorating progeny were considered for standardization, by using live-time anticoincidence counting (LTAC) in the Laboratório Nacional de Metrologia das Radiações Ionizantes (LNMRI) Brazil. The ionization chamber calibration factors were obtained using measurements of independent lots of 223Ra in LTAC considering the absolute method in order to evaluate the constancy and reproducibility of the standardization. The calibration factors of three high-performance well-type ionization chamber systems and four commercial well-type ionization chambers were determined. The LTAC results showed that lots 2 and 3 were 1069.88 and 1097.44 kBq/kg, with corresponding relative standard uncertainties 0.12% and 0.20%, respectively. The spectrometry method results of each lot value were classified as positive by ANOVA hypothesis testing. The emission probabilities relating to the X- and γ-rays in region from 81 to 830 keV ware determined. Hence, LNMRI can provide calibration services for 223Ra.