Investigating a CMUT's Ability to Achieve Non-linear Contrast Enhancement

PROCEEDINGS OF THE 2020 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS)(2020)

引用 0|浏览0
暂无评分
摘要
Capacitive micromachined ultrasonic transducers (CMUTs) are inherently non-linear and often believed to be unsuitable for non-linear contrast-enhanced ultrasound (CEUS) imaging. The ability of CMUTs to perform contrast-enhanced ultrasound imaging has been investigated by experimentally evaluating the contrast-to-noise ratio (CTR) and its variation with emitted pressure. This has been evaluated using amplitude modulation, pulse inversion and B-mode imaging sequences. The study shows that the CTR of the AM-mode was 7.5 dB higher than the B-mode on average, while the PI-mode is 5.9 dB lower than the B-mode. This indicates that contrast-enhancement can be achieved with a CMUT when a suitable image sequence is chosen.
更多
查看译文
关键词
Contrast-enhanced ultrasound, CEUS, capacitive micromachined ultrasound transducers, pulse inversion, amplitude modulation
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要