In infants and children, the small size of the systemic veins precludes safe and durable transvenous placement of automatic internal cardiac defibrillator (ICD). In such cases, an alternative method of implanting an automatic ICD is via thoracotomy. This report describes the technique used at Onassis Cardiac Surgery Center for the epicardial placement of an ICD system with a transvenous ICD coil placed in the posterior pericardium via a left thoracotomy, producing a stable location and excellent coil-to-can vector for successful defibrillation. To the best of our knowledge, this is the first case in Greece.