Ventricular tachycardia induced by pacing algorithm designed to avoid atrial fibrillation.

REVISTA PORTUGUESA DE CARDIOLOGIA(2020)

引用 0|浏览3
暂无评分
摘要
A patient with a dual chamber pacemaker was admitted to the emergency room after out-of-hospital cardioversion for syncopal sustained monomorphic ventricular tachycardia. Device interrogation revealed an abnormally timed ventricular spike after a ventricular premature beat at the beginning of the event, caused by a pacemaker algorithm designed to avoid atrial fibrillation, non-competitive atrial pacing. Despite the absence of significant coronary lesions, in the setting of a vulnerable substrate - a hypokinetic and hyperechogenic region of ventricular myocardium - an upgrade to a dual-chamber implantable cardioverter-defibrillator was performed, and substrate ablation was planned. (C) 2020 Sociedade Portuguesa de Cardiologia. Published by Elsevier Espana, S.L.U.
更多
查看译文
关键词
Ventricular tachycardia,Pacemaker algorithm,Non-competitive atrial pacing