Test-Retest-Reliability Of Computer-Based Metamorphopsia Measurement In Macular Diseases

KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE(2021)

引用 5|浏览16
暂无评分
摘要
In this study, the test-retest-reliability as one aspect of reliability of metamorphopsia measurements using a computer-based measuring method was determined in patients with macular diseases. Metamorphopsia amplitude, position, and area were quantified using AMD - A Metamorphopsia Detector software (app4eyes GmbH & Co. KG, Germany) in patients with diabetic, myopic, or uveitic macular edema, intermediate or neovascular age-associated macular degeneration, epiretinal membrane, vitelliform maculopathy, Irvine-Gass syndrome, or macular edema due to venous retinal occlusion. The intraclass correlation coefficient (ICC) was calculated in order to determine the repeatability of two repeated measurements and was used as an indicator of the reliability of the measurements. In this study, metamorphopsia measurements were conducted on 36 eyes with macular diseases. Metamorphopsia measurements made using AMD - A Metamorphopsia Detector software were highly reliable and repeatable in patients with maculopathies. The intraclass correlation coefficient of all indices was excellent (0.95-0.97). For diseases of the vitreoretinal interface or macular diseases with intra- or subretinal edema, this metamorphopsia measurement represents a supplement for visual function testing in the clinic, as well as in clinical studies.
更多
查看译文
关键词
meta morphopsia, quantitative metamorphopsia measurement, reliability, repeatability, Information Technology, test-retest-reliability