Technique for Removal of Broken InterTAN Integrated Cephalomedullary Screw

JOURNAL OF ORTHOPAEDIC TRAUMA(2021)

引用 1|浏览5
暂无评分
摘要
Pertrochanteric femoral fractures are routinely treated with cephalomedullary nailing, with good success. In the event of nonunion, implant fatigue failure may occur. The Trigen InterTAN system (Smith & Nephew, Memphis, TN) features 2 integrated cephalomedullary screws for improved rotational stability of the proximal segment. The inferior compression screw prevents rotation of the larger, superior lag screw. In the event of inferior screw breakage, the retained portion will prevent the integrated lag screw from being rotated to permit extraction. We present a case of a broken nail as well as a broken integrated screw, and describe a technique for successful implant removal.
更多
查看译文
关键词
intramedullary nail, interTAN, pertrochanteric fracture, broken implant, removal of implant, implant removal
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要